← Intona Answer Database (FAQ)
Can I use the 7055 series in a medical device? IEC 60601-1 and MOPP
The 7055-B and 7055-D are designed to meet 2 MOPP isolation requirements, the
level needed for IEC 60601-1 compliance in medical applications.
The isolator is a component. The final certification has to be done on your complete device, with the isolator installed.
To support your certification we can provide the bill of materials, CAD data,
isolation report and other documents free of charge. Contact support with the model
number and a short description of the application.